How-to • ESD compliance

ESD Testing Equipment: What to Test and How Often

An ESD control programme is only as good as its last test. Wrist strap cords break inside the insulation, mats lose their dissipative surface to the wrong cleaner, heel grounders wear through and ionisers drift, and none of it is visible. This guide sets out what to test in an Electrostatic Protected Area, with the limits from BS EN IEC 61340-5-1, how often to do it and what to write down.

Grove Sales • How-to

1.Why testing matters

Every ESD control is a resistance path to ground, and every one of them degrades. A wrist strap cord flexes thousands of times a week and the conductor eventually breaks inside an unmarked sleeve. Dissipative matting is wrecked by silicone polish or household cleaner. Footwear picks up an insulating layer of dirt. Because a failed control looks exactly like a working one, the standard requires compliance verification: a documented schedule of tests, with limits, instruments and records. The limits below are from BS EN IEC 61340-5-1; your own control plan may set tighter ones.

2.Daily: wrist straps

The wrist strap system, meaning the strap, its coiled cord and the person wearing it, must measure below 3.5 × 107 Ω. A tester checks that whole path at once: the wearer plugs in the cord and presses a metal plate, and the instrument passes or fails. Most testers pass between 750 kΩ and 35 MΩ; the lower limit catches a cord whose 1 MΩ safety resistor is missing or shorted, which would be a shock hazard.

For a small team a hand-held Wrist Strap Tester on the bench is enough. For an EPA with a controlled entrance, a wall-mounted Wrist Strap Test Station at the door lets everyone test and log before they go in.

3.Daily: footwear and heel grounders

Anyone who stands or moves around relies on ESD footwear or heel and toe grounders working with dissipative flooring. The footwear and flooring system must measure below 1 × 109 Ω, with the person's body voltage kept under 100 V, or below 3.5 × 107 Ω where body voltage is not measured. Footwear alone is tested standing on a metal footwear test plate connected to the tester, one foot at a time.

A combined station does both jobs in one unit and one log entry. The Complete Wrist Strap and Footwear Test Station tests either, and the Dual Footwear and Wrist Strap Test Station tests both feet and the strap at the same time, with a 100 MΩ footwear limit for control plans that apply the 1 × 108 Ω figure. Test footwear every day it is worn, and immediately after a change of flooring cleaner.

4.Continuous monitoring at the bench

A daily test proves the strap worked at 8 am. A constant monitor proves it is working now. It sits under or on the bench, the operator's cord plugs into it, and it watches the strap-to-person path and usually the bench mat ground as well, sounding an alarm and lighting an LED the moment either is lost. It also removes the daily strap test from the routine. For seated operators handling high-value or safety-critical boards, a monitor at every position is the norm.

The Constant Monitor uses capacitive sensing, so it works with a standard single-wire wrist strap rather than needing a dual-conductor one.

5.Periodic: work surfaces and flooring

Bench mats, work surfaces, shelving and flooring are tested with a surface resistance meter and a pair of 2.3 kg electrodes. Two measurements matter:

  • Resistance to groundable point (Rgp): from an electrode on the surface to the mat's earth stud or the bench's bonding point. Limit below 1 × 109 Ω for work surfaces and for flooring.
  • Point-to-point (Rpp): between two electrodes on the surface, which shows whether the surface itself is still dissipative or has been sealed by polish or dirt.

Dissipative mats normally read between 106 and 109 Ω. A rising trend is the early warning; a reading over the limit after cleaning with an ESD mat cleaner means replacement. Our Surface Resistance Meter does both tests, and the Surface Resistivity Kit adds the electrodes, leads and ground cord in a case so one kit covers mats, benches, flooring and racking.

6.Periodic: garments, packaging and ionisers

Garments. Check panel-to-panel resistance, sleeve to sleeve and sleeve to body, with the surface resistance meter: below 1 × 1011 Ω for a static control garment and below 1 × 109 Ω to the groundable point for a groundable one. Test on receipt, after repair and periodically after laundering. See Why Wear an ESD Lab Coat or Jacket?.

Packaging. Shielding bags and dissipative containers are specified to IEC 61340-5-3. Verify a sample of incoming packaging with the resistance meter and keep the supplier's certificates.

Ionisers. An ioniser that has drifted out of balance charges the work instead of neutralising it. Offset voltage should stay within ±35 V and decay time within your plan's limit. This needs a charged plate monitor, which we can supply to order; call us if you run ionisers.

7.How often, and what to record

ControlTestLimit (IEC 61340-5-1)How often
Wrist strap systemTester or test station< 3.5 × 107 ΩDaily before use, or constant monitor
Footwear and flooring systemFootwear tester on test plate< 1 × 109 Ω (with body voltage < 100 V) or < 3.5 × 107 ΩDaily when worn
Work surfaces and mattingSurface resistance meter, Rgp and Rpp< 1 × 109 ΩPer control plan, typically every 6 to 12 months and after a change of cleaner
FlooringSurface resistance meter< 1 × 109 ΩPer control plan, typically every 6 to 12 months
GarmentsPanel-to-panel resistance< 1 × 1011 Ω (groundable: < 1 × 109 Ω)On receipt, after repair, periodically after laundering
IonisersCharged plate monitorOffset within ±35 V; decay per planPer control plan, typically every 6 months

Record the date, the item, the reading, the instrument and the operator, and what you did about any failure. The records are what an ISO 9001 or customer auditor reads first, and they are what turns a box of testers into a control programme.

Frequently asked questions

How often should ESD wrist straps be tested?

Every day, before the wearer starts handling sensitive items, and again if the strap is removed and refitted. Many sites test at the EPA entrance with a wall-mounted test station and log the result. The alternative is a constant monitor at the bench, which checks the strap continuously and removes the need for a daily test.

What resistance should an ESD mat or work surface read?

To IEC 61340-5-1 the resistance from any point on the surface to its groundable point must be below 1 × 10⁹ Ω. Most dissipative matting reads between 10⁶ and 10⁹ Ω. A reading that has climbed towards the limit usually means the mat is dirty or has been cleaned with the wrong product, so clean it with an ESD mat cleaner and re-test before replacing it.

Do I need a constant monitor?

Not under the standard, which accepts a daily test, but it is the only way to know a strap has not failed between tests. Cords break inside the insulation, poppers loosen and people unplug to reach something. For high-value or safety-critical assemblies a monitor at each seated position is the usual choice.

What records should I keep?

Daily wrist strap and footwear results with the operator's name, periodic surface, flooring, garment and ioniser readings with the instrument used, and a note of any failure and what was done about it. Auditors for ISO 9001 and for your customers will ask for them, and a gap in the log looks like a gap in the control.